TB 9-6695-288-35
Test instrument
Calibrator indications
Meter INPUT
Meter
RANGE switch
indications
Min
Max
settings
.1 mV1
1
10 Hz
.098 V
.102 V
10 kHz
.098 V
.102 V
.3 mV1
3
10 Hz
.288 V
.312 V
10 kHz
.294 V
.306 V
.001 V
1
20 Hz
0.98 V
1.02 V
10 kHz
0.98 V
1.02 V
.003 V
3
20 Hz
2.88 V
3.12 V
10 kHz
2.94 V
3.06 V
010 V2
1
20 Hz
9.6 mV
10.4 V
10 kHz
9.8 mV
10.2 mV
100 kHz
9.6 mV
10.4 mV
.030 V
3
20 Hz
28.8 mV
31.2 mV
10 kHz
29.4 mV
30.6 mV
100 kHz
28.8 mV
31.2 mV
.1 V
1
20 Hz
96.0 mV
104.0 mV
10 kHz
98.0 mV
102.0 mV
100 kHz
96.0 mV
104.0 mV
.3 V
3
20 Hz
288 mV
312 mV
10 kHz
294 mV
206 mV
100 kHz
288 mV
312 mV
1V
1
20 Hz
0.96 V
1.04 V
1
10 kHz
0.98 V
1.02 V
1
100 kHz
0.96 V
1.04 V
0.8
1 kHz
0.78 V
0.82 V
0.6
1 kHz
0.58 V
0.62 V
0.4
1 kHz
0.38 V
0.42 V
0.2
1 kHz
0.18 V
0.22 V
3V
3
20 Hz
2.88 V
3.12 V
10 kHz
2.94 V
3.06 V
100 kHz
2.88 V
3.12 V
10 V
1
20 Hz
9.6 0 V
10.4 V
10 kHz
9.80 V
10.2 V
100 kHz
9.60 V
10.4 V
30 V
3
20 Hz
28.8 V
31.2 V
10 kHz
29.4 V
30.6 V
100 kHz
28.8 V
31.2 V
100 V
1
20 Hz
96 V
104 V
10 kHz
98 V
102 V
100 kHz
96 V
104 V
300 V
3
50 Hz
288 V
312 V
10 kHz
294 V
306 V
100 kHz
288 V
312 V
1For
Option 001 only.
2Remove
ratio transformer from equipment setup for this and remaining checks.
6 CHANGE 1