TB 9-6695-293-50
Test Instrument
Performance specifications
parameters
Calibrator, John Fluke, Model 5700A/( ) and 5720A/( )
Dc current
Output: 0 to 2.2 A
Accuracy: (ppm reading + nA)
A.....
Range:
220
60 +
10
2.2
mA.....
60 +
10
22
mA.....
60 +
100
A
220
m A.....
70 +
1.0
A
2.2
A.....
95 +
30
Ac current
Output: 9 A to 2.2 A
Accuracy: (ppm reading + nA)
Range:
A
220
10
20 Hz
800 +
30
20
40 Hz
420 +
25
40 Hz 1.0 kHz
160 +
20
1.0
5.0 kHz
700 +
50
5.0
10 kHz
.18 % +
100
2.2
mA
10
20 Hz
800 +
50
20
40 Hz
420 +
40
40 Hz 1.0 kHz
160 +
40
1.0
5.0 kHz
700 +
500
A
5.0 10 kHz
.18 % +
1.0
22
mA
10 20 Hz
800 +
500
20 40 Hz
420 +
400
40 Hz 1.0 kHz
160 +
400
A
1.0 5.0 kHz
700 +
5.0
A
5.0 10 kHz
.18 % +
10
A
220
mA
10 20 Hz
800 +
5
A
20 40 Hz
420 +
4
A
40 Hz 1.0 kHz
180 +
4
A
1.0 5.0 kHz
700 +
50
A
5.0 10 kHz
.18 % +
100
A
2.2
A
20 Hz 1.0 kHz
750 +
40
A
1.0 5.0 kHz
850 +
100
A
5.0 10 kHz
1.0 % +
200
Ac voltage
Output: 220 V 1100 V
Accuracy: (ppm reading + V)2
Range:
2.2
mV
10
20 Hz
600 +
5
20
40 Hz
240 +
5
40 Hz 20 kHz
120 +
5
20
50 kHz
410 +
5
50
100 kHz
950 +
8
100
300 kHz
.13 % +
15
300
500 kHz
.18 % +
30
500 kHz 1.0 MHz
.48 % +
40
22
mV
10
20 Hz
600 +
6
20
40 Hz
240 +
6
40 Hz 20 kHz
120 +
6
20
50 kHz
410 +
6
50
100 kHz
950 +
8
See footnotes at end of table.