TB 9-6695-293-50
2. Forms,
Records, and
Reports
a.
Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. No physical adjustments will be performed on the calibrator, Model 5700A/( ),
5720A/( ) or amplifier, Model 5725A/( ). Instead, an artifact calibration will be performed
prior to performance verification tests. If an out-of-tolerance condition is detected, verify
condition of standards, accessories, and equipment connections. If TI indications remain
out of tolerance, repair action is required.
c. Adjustments to models 5215A/CT and 5220A/CT should be recorded and are
designated (R) at the end of the sentence in which they appear. Report only those
adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
parameters
Performance specifications
Calibrator, John Fluke, Model 5700A/( ) and 5720A/( )
Dc voltage
Output: 0 to 1100 V
Accuracy: (ppm reading + V)
Range :
220
mV.....
9+
0.8
2.2
V.....
8+
1.2
11
V.....
8+
4
22
V.....
8+
8
220
V.....
9 + 100
1100
V.....
11 + 600
Output: 0 to 100 MΩ
Accuracy: (ppm ) 1
Nominal output:
0 Ω
50 Ω
110
1.0 Ω
110
1.9 Ω
33
10 Ω
31
19 Ω
....
20
100 Ω
20
190 Ω
15
1.0 kΩ
15
1.9 kΩ
14
10 kΩ ....
14
19 kΩ
16
100 kΩ
16
190 kΩ
23
1.0 MΩ
24
1.9 MΩ ....
46
10 MΩ
55
19 MΩ
130
100 MΩ
See footnotes at end of table.