TB 9-6695-294-35
NOTE
When the CALIBRATION switch is set to ON, some of the
pushbuttons on the front panel are used to enter and store
calibrator output display indication of 9.999467 in b(13)
below, press STORE pushbutton, OHMS Ω (9) pushbutton,
MANUAL/EXTERNAL (.) pushbutton, OHMS Ω (9)
pushbutton thrice, AUTO (4) pushbutton, RANGE ↓ (6)
pushbutton, and NORMAL (7) pushbutton.
(13) Set calibrator for a 10 Ω output. Allow TI to take at least one measurement, then press
STORE (calibrator output display indication) (CAL COR) (R).
(14) Set calibrator for a 100 Ω output. Allow TI to take at least one measurement, then press
STORE (calibrator output display indication) (CAL COR) (R). Example: STORE (99.9987) (CAL
COR)
(15) Set calibrator for a 1 kΩ output. Allow TI to take at least one measurement, then press
STORE (calibrator output display indication) (CAL COR) (R). Example: STORE (1000.003) (CAL
COR)
(16) Set calibrator for a 10 kΩ output. Allow TI to take at least one measurement, then press
STORE (calibrator output display indication) (CAL COR) (R). Example: STORE (9999.95) (CAL
COR)
(17) Set calibrator for a 100 kΩ output. Allow TI to take at least one measurement, then press
STORE (calibrator output display indication) (CAL COR) (R). Example: STORE (99997.9) (CAL
COR)
(18) Set calibrator for a 1 MΩ output. Allow TI to take at least one measurement, then press
STORE (calibrator output display indication) (CAL COR) (R). Example: STORE (1.000004 EXP 6)
(CAL COR)
(19) Set calibrator for a 10 MΩ output. Allow TI to take at least one measurement, then press
STORE (calibrator output display indication) (CAL COR) (R). Example: STORE (9.99699 EXP 6)
(CAL COR)
11. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA Label/Form in accordance with TB 750-25.
12