TB 9-6695-294-35
NOTE
Ac software corrections are only made at 1 kHz. If an out-of-
tolerance condition exists in (9) below or in table 5, perform
the software correction in b below which corresponds to the
range that is out of tolerance, using the 1 kHz technique of (2)
through (4) above and/or table 4. If the software correction
had previously been completed for the range in question,
repair action may be required.
(7) Press TI RANGE ↑ or ↓ pushbutton to select the 100 mV range.
(8) Set calibrator for a 100 mV, 30 Hz output.
(9) Allow TI to take several measurements. TI will indicate between 99.890 (-3) and 100.110
(-3).
(10) Repeat technique of (7) through (9) above using settings listed in table 5. TI will indicate
within the specified limits.
CAUTION
Do not exceed 600 V ac with TI in AUTO RANGE.
Table 5. Ac Voltage
Calibrator output
Test instrument
Test instrument
settings
indications
range settings
Voltage
Min
Max
100 mV
100 mV
10 kHz
99.835 (-3)
100.165 (-3)
100 mV
100 mV
30 kHz
99.710 (-3)
100.290 (-3)
100 mV
100 mV
90 kHz
99.710 (-3)
100.290 (-3)
300 mV
300 mV
30 Hz
299.688 (-3)
300.312 (-3)
300 mV
300 mV
10 kHz
299.730 (-3)
300.270 (-3)
300 mV
300 mV
30 kHz
299.280 (-3)
300.720 (-3)
300 mV
300 mV
90 kHz
299.280 (-3)
300.720 (-3)
1
V
1
V
30 Hz
.99896
1.00104
1
V
1
V
10 kHz
.99910
1.00090
1
V
1
V
30 kHz
.99760
1.00240
1
V
1
V
90 kHz
.99760
1.00240
3
V
3
V
30 Hz
2.99688
3.00312
3
V
3
V
10 kHz
2.99730
3.00270
3
V
3
V
30 kHz
2.99280
3.00720
3
V
3
V
90 kHz
2.99280
3.00720
10 V
10 V
30 Hz
9.9896
10.0104
10 V
10 V
10 kHz
9.9910
10.0090
10 V
10 V
30 kHz
9.9760
10.0240
10 V
10 V
90 kHz
9.9760
10.0240
30 V
30 V
30 Hz
29.9688
30.0312
30 V
30 V
10 kHz
29.9730
30.0270
30 V
30 V
30 kHz
29.9280
30.0720
30 V
30 V
90 kHz
29.9280
30.0720
8