TB 9-6695-294-35
(3) Press TI RANGE ↓ pushbutton to select 100 mV range.
(4) Short TI VOLTS INPUT/OHMS SENSE HI and LO terminals using a
good quality, low-thermal shorting bar. Allow sufficient thermal settling time before
proceeding.
(5) Press ZERO V DC/Ω pushbutton (ZERO annunciator off).
(6) Allow TI to take at least one reading then press ZERO V DC/Ω pushbutton (ZERO
annunciator on). This operation stores new ZERO correction factors.
(7) Press RANGE ↑ pushbutton and repeat (5) and (6) above.
(8) Repeat (7) above for remaining ranges.
(9) Disconnect shorting bar from TI.
(10) Connect calibrator OUTPUT HI, LO and V-GUARD to TI VOLTS INPUT/OHMS
SENSE HI, LO and GD.
(11) Press TI RANGE ↓ pushbutton to select 10 V range.
NOTE
pushbuttons on the front panel are used to enter and store
new numeric calibration constants. Example: to enter the
10.000000 in b(12) below, you would press the EXT REF (1)
pushbutton, LCL/RMT (0) pushbutton, MANUAL/EXTERNAL (.)
pushbutton and then the LCL/RMT (0) pushbutton six
times.
(12) Set calibrator for a 10 V dc output. Allow TI to take at least one measurement, then press
STORE (10.000000) (CAL COR).
(13) Set calibrator for a 100 mV dc output.
(14) Press TI RANGE ↓ pushbutton to select 100 mV range.
(15) Allow TI to take at least one measurement, then press STORE (100.00000 EXP-3)
(CAL COR).
(16) Press TI RANGE ↑ pushbutton to select 1 V range.
(17) Set calibrator for a 1 V dc output. Allow TI to take at least one measurement, then press
STORE (1.000000) (CAL COR).
(18) Press TI RANGE ↑ pushbutton to select 100 V range.
(19) Set calibrator for a 100 V dc output. Allow TI to take at least one measurement, then
press STORE (100.0000) (CAL COR).
(20) Press TI RANGE ↑ pushbutton to select 1000 V range.
6