TB 9-6695-294-35
SECTION I
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification. This
bulletin provides instructions for the calibration of
Digital
Multimeter, John Fluke, Model 8506A/CT. The manufacturer's manual and MIS-35947 were used as
the prime data sources in compiling these instructions. The equipment being calibrated will be referred
to as the TI (test instrument) throughout this bulletin.
a. Model Variations . None.
b. Time and Technique . The time required for this calibration is approximately 2 hours, using the
2. Forms, Records, and Reports
a. Forms,
records, and reports required for calibration personnel at
all levels are prescribed by TB
750-25.
b. Software corrections to be reported are designated (R) at the end of the sentence in which they
appear. Report only those software corrections made and designated with (R). Use the range or
parameter being corrected to report software corrections; i.e., V dc, 100 V ac.
3. Calibration Description. TI parameters and performance specifications which pertain to this
Table 1. Calibration Description
Test instrument parameters
Performance specifications 1
Dc voltage
Range: 0 to 1000 V
Range
Accuracy: (% of reading + % of range)
100
.004 +.001
1 V through 1000 V .005 + 0
Ac voltage
Range: 0 to 1000 V rms (in 9 ranges)
Accuracy:
(% of reading + % of range)
Frequency
Range
100 mV, 500 V,
300 mV
and 1000 V
through 100 V
20
to 50
.1 + .01
.1 + .004
Hz
50 Hz
to
2 kHz
.05 + .015
.04 + .01
2 to
20
kHz
.14 + .025
.07 + .02
20
to
100 kHz
.24 + .05
.2 + .04
See footnotes at end of table.
2