TB 9-6695-299-50
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of this calibration.
DEATH ON CONTACT may result if personnel fail to observe safety precautions.
REDUCE OUTPUT(S) to minimum after each step within the performance check
where applicable.
a. Utilizing the dc
reference standard and resistance standard, characterize multimeter to
obtain 24-hour
manufacturer specifications.
NOTE
If ambient temperature drifts more than 1 degree Celsius prior to completing
b. Configure multimeter as
listed in
(1) through (6) below:
(1) DCV.
(2) RANGE AUTO.
(3) NDIG 8.
(4) NLPC 200.
(5) Trig SGL.
(6) Guard pushbutton to Open (released).
c.
Connect high quality test leads to TI as
illustrated in
figure 1.
4