TB 9-6695-299-50
9. Calibration Test Report
a. A calibration test report is required for this TI.
b. Calibration activities will maintain a
case history file of
test reports for the TI. The file will contain an
accumulation
of at least six of the most recent test reports.
c. An
example of
calibration data to be
annotated on
the test report is in
appendix B.
d. Transcribe V
IN and multimeter indication data from table 3, sections A and B, to calibration test report.
e. Calculate, and record, the mV/V value for the calibration test report, using the formula:
(VOUT/ VIN) X 1000 = mV/V output
Where:
VOUT = recorded multimeter indication from front terminals in volts
VIN = recorded multimeter indication from rear terminals in volts
10. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
8