TB 9-4920-460-24
(10) Repeat technique of (9) above for remaining settings listed in table 5. If
measuring receiver does not indicate within limits specified in table 5, perform b (3)
through (21) below.
Table 5. Transmitter Power RT.
Measuring receiver
Test instrument
indication (dBm)
RF OUTPUT
FUNCTION
LEVEL
(manually
Min
Max
switch setting
entered) (MHz)
(dBm)
MB
75
-35
-37.0
-33.0
MB
75
-75
-77.5
-72.5
MB
75
-90
-93.0
-87.0
VOR
112.05
-40
-42.0
-38.0
VOR
112.05
-70
-72.0
-68.0
VOR
112.05
-80
-82.0
-78.0
LOC
110.10
-55
-57.0
-53.0
LOC
110.10
-85
-87.0
-83.0
LOC
110.10
-95
-97.0
-93.0
GS
108.95
-35
-37.0
-33.0
GS
108.95
-55
-57.0
-53.0
GS
108.95
-75
-78.0
-72.0
GS
110.15
-40
-42.0
-38.0
GS
110.15
-60
-62.0
-58.0
GS
110.15
-80
-83.0
-77.0
29-88
46
-30
-32.0
-28.0
29-88
46
-60
-62.0
-58.0
29-88
46
-90
-92.0
-88.0
29-88
63
-40
-42.0
-38.0
29-88
63
-70
-72.0
-68-0
29-88
63
-110
-113.0
-107.0
118-156
135
-35
-37.0
-33.0
118-156
135
-50
-52.0
-48.0
118-156
135
-65
-67.0
-63.0
118-156
150
-40
-42.0
-38.0
118-156
150
-55
-57.0
-53.0
118-156
150
-70
-72.0
-68.0
156-174
160
-45
-47.0
-43.0
156-174
160
-60
-62.0
-58.0
156-174
160
-75
-77.0
-73.0
225-400
250
-30
-32.0
-28.0
225-400
250
-50
-52.0
-48.0
225-400
250
-65
-67.0
-63.0
225-400
308
-40
-42.0
-38.0
225-400
308
-60
-62.0
-58.0
225-400
308
-75
-78.0
-73.0
225-400
366
-35
-37.0
-33.0
225-400
366
-55
-57.0
-53.0
225-400
366
-75
-78.0
-72.0
225-400
395
-30
-32.0
-28.0
225-400
395
-45
-47.0
-43.0
225-400
395
-65
-67.0
-63.0
(11) Disconnect equipment setup and turn TI off.