TB 9-4931-537-35
(2) This function takes approximately 13 minutes to run. While waiting, and
with nothing connected to Output terminals, perform DC Zero adjustment on calibrator
by pressing Setup Menus , Cal, and Zero softkeys
d. Dc Voltage. Calibrator, John Fluke, Model 5700A/( ) and Multimeter, Hewlett
Packard, Model 3458A.
CALIBRATOR
OUTPUT
SENSE
HI
HI
MULTIMETER
LO
LO
SENSE INPUT
HI
HI
HI
GUARD GROUND
LO
LO
CURRENT
GUARD
I
OUTPUT
HI
LO
AMPLIFIER
(2) Set multimeter Guard switch to Open. Configure multimeter NDIG to 8 (8
digit display) by pressing the following pushbuttons: FUNCTION/RANGE, blue colored
key, MENU N, then NUMERIC/USER 8 followed by Enter.
(3) Configure multimeter NPLC to 100 (100 power line cycles) by pressing the
following pushbuttons: MENU NPLC then NUMERIC/USER 1, 0, 0, followed by
Enter.
(4) Set calibrator to (m)V dc outputs listed in table 1.
Record multimeter
indications in space produced.
6