TB 9-6625-116-24
Table 3. DC Voltage Accuracy Continued.
TI
Output (V dc)
Min
Measured
Max
0.06
59.96000
mV
60.04000 mV
-0.06
-60.04000
mV
-59.96000 mV
0.19
0.1899275
V
0.1900725 V
-0.19
-0.1900725
V
-0.1899275 V
0.23
0.2299175
V
0.2300825 V
-0.23
-0.2300825
V
-0.2299175 V
0.5
0.4998500
V
0.5001500 V
-0.5
-0.5001500
V
-0.4998500 V
0.6
0.5998250
V
0.6001750 V
-0.6
-0.6001750
V
-0.5998250 V
1.9
1.899500
V
1.900500 V
-1.9
-1.900500
V
-1.899500 V
2.3
2.299400
V
2.300600 V
-2.3
-2.300600
V
-2.299400 V
5
4.998725
V
5.001275 V
-5
-5.001275
V
-4.998725 V
6
5.998475
V
6.001525 V
-6
-6.001525
V
-5.998475 V
19
18.99523
V
19.00478
V
-19
-19.00478
V
-18.99523
V
23
22.99423
V
23.00578
V
-23
-23.00578
V
-22.99423
V
50
49.98748
V
50.01253
V
-50
-50.01253
V
-49.98748
V
60
59.98498
V
60.01503
V
-60
-60.01503
V
-59.98498
V
190
189.9525
V
190.0475
V
-190
-190.0475
V
-189.9525
V
(6) Set TI OUTPUT to OFF.
(7) Set multimeter to measure ACV, SETACV to SYNC mode.
(8) Set TI as listed in (a) through (e) below:
(a) [WAVE FORM]
(b) [
] (positive square wave)
(c) Keypad Entry, 0.006
(d) [V]
(e) OUTPUT, ON
(9) Multimeter will indicate within the limits of table 4. If not, perform b below.
6