TB 9-6625-133-40
TI
Nominal
Test report
Range
Mode
Indication
Min
Measured
Max
(Ω)
(Ω)
(Ω)
(Ω)
(Ω)
(ppm)
(ppm)
(ppm)
1
2
Normal
-27.2
27.2
1
2
LoI
-27.2
27.2
1
2
Tru
-27.2
27.2
1
2
Tru LoI
-27.2
27.2
10
20
Normal
-14.0
14.0
10
20
LoI
-14.0
14.0
10
20
Tru
-14.0
14.0
10
20
Tru LoI
-14.0
14.0
100
200
Normal
-10.8
10.8
100
200
LoI
-12.0
12.0
100
200
Tru
-10.8
10.8
100
200
Tru LoI
-12.0
12.0
1k
2k
Normal
-10.8
10.8
1k
2k
LoI
-12.0
12.0
1k
2k
Tru
-10.8
10.8
1k
2k
Tru LoI
-12.0
12.0
10 k
20 k
Normal
-10.8
10.8
10 k
20 k
LoI
-12.0
12.0
10 k
20 k
Tru
-10.8
10.8
10 k
20 k
Tru LoI
-12.0
12.0
100 k
200 k
Normal
-10.8
10.8
100 k
200 k
LoI
-11.4
11.4
1M
2M
Normal
-13.4
13.4
1M
2M
LoI
-18.4
18.4
10 M
20 M
Normal
-37.2
37.2
10 M
20 M
LoI
-122.2
122.2
HiVΩ
10 M1
20 M
-21.4
21.4
100 M2
200 M
Normal
-270.2
270.2
100 M
200 M
LoI
-3010.2
3010.2
HiVΩ
100 M1
200 M
-92.2
92.2
1G
2G
Normal
-3010.2
3010.2
1G
2G
LoI
-3010.2
3010.2
HiVΩ
1 G1
2G
-350.2
350.2
To avoid electrical shock, press Ω to disable HiVΩ when measurement is complete, then make connection changes.
1
Disable 4-wire ohms mode for all measurements > 20 MΩ.
2
16