TB 9-6625-154-40
(17) Set TI as listed in (a) through (g) below:
(a)
Softkey, EXIT.
(b)
Softkey, STORE HEAD CAL.
(c)
Softkey, STORE.
(d)
Wait for Saving Cal factors to head to complete.
(e)
Softkey, EXIT twice.
(f)
MODE.
(g)
Softkey, MANUAL.
(18) Set TI output to OFF, disconnect storage oscilloscope and repeat paragraph 12 a
above.
a. Store new due date to TI by setting 9500B (base) as listed in (1) through (13) below:
(1) Rear-panel CAL ENABLE/DISABLE switch to ENABLE.
(2) MODE.
(3) Softkey, CALIB.
(4) Enter Password (exp. 2→3→5→7 ).
(5) Softkey, HEAD CAL.
(6) Softkey, STORE HEAD CAL.
(7) Enter new due date using keypad, then press .
(8) Select warning period, as necessary, using Softkey (7-Day recommended).
(9) Softkey, STORE.
(10)Softkey, EXIT twice.
(11)MODE.
(12)Softkey, MANUAL.
(13)Rear-panel CAL ENABLE/DISABLE switch to DISABLE.
b. De-energize and disconnect all equipment.
c. Annotate and affix DA label 163 in accordance with TB 750-25. For model 9530
only, annotate "Capacitance measure function not calibrated."
Change 1