TB 9-6625-197-24
IDENTIFICATION AND DESCRIPTION
This bulletin provides instructions for the
1. Test Instrument Identification.
The manufacturer's manuals were used as the prime data source in compiling these
instructions. The equipment being calibrated will be referred to as the TI (test instrument)
throughout this bulletin.
Variations among models are described in text, tables and
a. Model Variations.
figures.
b. Time and Technique. The time required for this calibration is approximately 3
hours, using the dc and low frequency techniques.
2. Forms, Records, and Reports. Forms, records, and reports required for calibration
personnel at all levels are prescribed by TB 750-25.
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
DC calibrator
Range: 5 V dc; 0.00 V
Accuracy: 10 mV; 500 μV
DC volts dual cursor
Range: 2 mV to 5 V/div; (0.1 V and 0.5 V/div for CH 3 & 4, 54601A &
54602A)
Accuracy: (1.5% rdg + 0.4% FS)1
Range: DC to 100 MHz, (54600A & 54601A); DC to 150 MHz,
(54602A);
CH 3 & 4: DC to 100 MHz, (54601A); DC to 250 MHz, (54602A)
Accuracy: down not more than 3 dB
Time measurement2
Range: 2 ns to 5 s/div
Accuracy: (0.01% rdg + 0.2% FS + 200 ps)
Triggering
Range: (all CH) 0.35 div or 3.5 mV p-p, DC to 25 MHz;
INT:
(54600A & 54601A, all CH) 1 div or 10 mV p-p, DC to 100 MHz;
(54602A, CH 1 & 2) 1 div or 10 mV p-p, DC to 150 MHz;
(54602A, CH 3 & 4) 1 div or 100 mV p-p, DC to 250 MHz
Accuracy: minimum
EXT:
Range: (54600A only): 50 mV, DC to 25 MHz;
100 mV, DC to 100 MHz
Accuracy: minimum
FS is 80 mV for 2 mV and 5 mV Volts/Div ranges.
1
Specification applies to Ćt, 1/Ćt and time cursor measurements.
2