TB 9-6625-2043-24
(4) Set FULL SCALE switch to -30 dBm.
(5) Adjust function/arbitrary generator rms output for an indication of -3.0 V dc on
multimeter. If true rms voltmeter does not indicate between 7.04 and 7.1 mV
a c , p e r f o r m b below.
b. Adjustments
(1) Adjust function/arbitrary generator output for an indication of 7.07 mV on true
(2) Adjust R180 (fig. 1 or 2) for a -3.0 V dc indication on multimeter (R).
11. Recorder Full Scale Sensitivity
a. Performance Check
(1) Disconnect positive side of multimeter from TP17 and connect to + (positive)
RECORDER connector on TI rear panel.
(2) Zero TI meter as described in paragraph 7 h above.
(3) Set FULL SCALE switch to -30 dBm.
(4) Adjust function/arbitrary generator output for an indication of 10.0 V dc on
multimeter. If true rms voltmeter does not indicate between 7.04 and 7.1 mV, perform b below.