TB 9-6625-2049-24
Test instrument settings
Multimeter indications
ENTRY AMPTD
ENTRY FREQ
(V ac)
3.536
V RMS
1 kHz
3.495
3.577
3.536
V RMS
100 kHz
3.495
3.577
1.061
V RMS
100 kHz
1.048
1.073
1.061
V RMS
1 kHz
1.048
1.073
1.061
V RMS
100 Hz
1.048
1.073
.3536 V RMS1
100 Hz
0.3416
0.3660
.3536 V RMS
1 kHz
0.3416
0.3660
.3536 V RMS
100 kHz
0.3416
0.3660
1
Press ENTRY DC OFFSET key and enter 1 mV using DATA keys.
(5) Press keys and enter values using DATA keys as listed in (a) through (g) below:
ENTRY DC OFFSET.
(a)
0 mV.
(b)
ENTRY FREQ.
(c)
1 kHz.
(d)
ENTRY AMPTD.
(e)
.707 V RMS.
(f)
AMPTD CAL.
(g)
(6) Press MODIFY keys for a multimeter indication as close as possible to 0.707 V ac.
(7) Enter 100 kHz using DATA keys. Multimeter will indicate between 0.683 and
0.732 V ac. Record multimeter indication.
(8) Disconnect multimeter and 50
feedthrough termination from TI.
NOTE
Zero and calibrate measuring receiver sensor module.
(9) Connect measuring receiver sensor module to TI SIGNAL.
(10) Set measuring receiver to measure power in volts at .1 MHz.
(11) Press ENTRY AMPTD key.
(12) Press MODIFY keys for a measuring receiver indication as close as possible to
indication recorded in (7) above.
(13) Press ENTRY FREQ key and enter 500 kHz using DATA keys.
(14) Set measuring receiver to measure power in volts at .5 MHz.
Measuring
receiver will indicate between 0.683 and 0.732 V.
(15) Repeat technique of (13) and (14) above using TI and measuring receiver
settings listed in table 4. Measuring receiver will indicate between 0.683 and 0.732 V.
15