TB 9-6625-2049-24
(11) Enter -20 VOLT using DATA keys. Multimeter will indicate between -19.775
and -20.225 V dc.
b. Adjustments. No adjustments can be made.
a. Performance Check
f
(1) Connect SIGNAL to multimeter INPUT HI and LO using 50
eedthrough termination.
(2) Press keys and enter values using DATA keys as listed in (a) through (i) below:
SIGNAL off (option 002 not lit).
(a)
FUNCTION sine wave.
(b)
ENTRY FREQ.
(c)
20.999 999 999 MHz.
(d)
ENTRY AMPTD.
(e)
1 VOLT.
(f)
ENTRY DC OFFSET.
(g)
4.5 VOLT.
(h)
AMPTD CAL.
(i)
(3) Multimeter will indicate between 4.35 and 4.65 V dc.
(4) Enter -4.5 VOLT using DATA keys. Multimeter will indicate between -4.35 and
-4.65 V dc.
(5) Press ENTRY FREQ key and enter 999.9 kHz using DATA keys. Multimeter
will indicate between -4.44 and -4.56 V dc.
(6) Press ENTRY DC OFFSET key and enter 4.5 VOLT using DATA keys.
Multimeter will indicate between 4.44 and 4.56 V dc.
(7) Press FUNCTION square wave key. Multimeter will indicate between 4.44 and
4.56 V dc.
(8) Enter -4.5 VOLT using DATA keys. Multimeter will indicate between -4.44 and
-4.56 V dc.
(9) Press ENTRY FREQ key and enter 9.9999 MHz using DATA keys.
Multimeter will indicate between -4.35 and -4.65 V dc.