TB 9-6625-2059-35
a. Performance Check
(1) Connect signal generator RF output to power meter.
(2) Adjust signal generator for 100 MHz and amplitude for 0 dBm indication on
power meter.
(4) Set step attenuator to 80 dB.
(5) Without adjusting signal generator, establish a reference indication on
spectrum analyzer.
(6) Repeat technique of (1) through (5) above, for power meter reading of -2.5 and
+2.5 dBm. Record reference on spectrum analyzer without changing setting on
spectrum analyzer.
SIGNAL GENERATOR
SPECTRUM ANALYZER
STEP
ATTENUATOR