TB 9-6625-2155-35
(7) Verify pulse overshoot and undershoot is less than 20% and ALC UNLEVEL
light on TI is extinguished. If signal is not within limits specified perform adjustments
listed in b below.
(8) Set TI, and local oscillator to the remaining frequencies and levels listed in table
14, and
repeat (7) above at
each frequency.
Table 14. Pulse Modulation Overshoot and Undershoot Check
Test instrument
LO
Level
(GHz)
(dB)
(GHz)
Description
2.00
3
2.050
Overshoot Undershoot @ 2.0 GHz
+3dBm
2.00
-10
2.050
Overshoot Undershoot @ 2.0 GHz
-10dBm
6.60
3
6.650
Overshoot Undershoot @ 6.6 GHz
+3dBm
6.60
-10
6.650
Overshoot Undershoot @ 6.6 GHz
-10dBm
6.60
0
6.650
Overshoot Undershoot @ 6.6 GHz
0dBm
6.70
3
6.750
Overshoot Undershoot @ 6.7 GHz
+3dBm
6.70
-10
6.750
Overshoot Undershoot @ 6.7 GHz
-10dBm
6.70
0
6.750
Overshoot Undershoot @ 6.7 GHz
0dBm
12.290
3
12.340
Overshoot Undershoot @ 12.29 GHz
+3dBm
12.290
-10
12.340
Overshoot Undershoot @ 12.29 GHz
-10dBm
20