TB 9-6625-2239-24
Figure 3. Tektronix, Type 465B with options top view.
(3) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton to
adjust output for 0.0% indication on err display. Adjust R4443 (fig. 1) for 6 divisions of
vertical display (R).
(4) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton to
adjust output for 0.0% indication on err display. Adjust R4272 (fig. 1) for 6 divisions of
vertical display (R).
9