Frequency Span Accuracy - Continued
Frequency Span Accuracy - Continued

Calibration Procedure For Spectrum Analyzer An/Usm-489A Page Navigation
  23    24    25    26    27  28  29    30    31    32    33  
TB 9-6625-2250-24
(30) Press TI keys as listed in (a) through (c) below:
(a) CONTROL TRIG.
(b) [SINGLE].
(c) MARKER PEAK SEARCH.
MKR indication will be between 38 and 42 MHz.
(31) TI
(32) Press TI SPAN key then [SPAN WIDTH] to 5 GHz.
(33) Set synthesized signal generator frequency to 7 GHz.
(34) Press TI keys as listed in (a) through (e) below:
MARKER OFF.
(a)
CONTROL TRIG.
(b)
[SINGLE].
(c)
MARKER PEAK SEARCH.
(d)
[MARKER DELTA].
(e)
(35) Set synthesized signal generator frequency to 11 GHz.
(36) Press TI keys as listed in (a) through (c) below:
(a) CONTROL TRIG.
(b) [SINGLE].
(c) MARKER PEAK SEARCH.
MKR indication will be between 3.8 and 4.2 GHz.
(37) TI
(38) Press TI FREQUENCY key then [CENTER FREQ] to 16 GHz.
(39) Set synthesized signal generator frequency to 16 GHz.
(40) Press TI keys as listed in (a) through (g) below:
MARKER OFF.
(a)
CONTROL TRIG.
(b)
[CONT].
(c)
MARKER PEAK SEARCH.
(d)
INSTRUMENT STATE MIXER INT.
(e)
[PRESEL AUTO PK] wait for PEAKING message to disappear.
(f)
SPAN then [SPAN WIDTH] to 10 kHz.
(g)
(41) Set synthesized signal generator frequency to 15.999996 GHz.
(42) Press TI keys as listed in (a) through (e) below:
MARKER OFF.
(a)
CONTROL TRIG.
(b)
[SINGLE] wait for completion of sweep.
(c)
MARKER PEAK SEARCH.
(d)
31