TB 9-6625-2262-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
calibration of Multimeter Fluke, Model 8025A and APN 13235191 (Fluke, Model 8025B).
The manufacturer's manual and specification control drawing No. 13235191 were used as
the prime data sources in compiling these instructions. The equipment being calibrated
will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. Variations among models are described in text.
b. Time and Technique. The time required for this calibration is approximately 1
hour, using the dc and low frequency technique.
2. Forms,
Records, and
Reports
a.
Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be recorded are designated (R) at the end of the sentence in which
they appear. Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
Performance specifications
parameters
(% of reading + digits) (31/2 digit display)
Dc voltage
Range: 0 to 1000 V
Accuracy: 0.2 + 1
Accuracy
Ac voltage
Range
40 Hz to 2 kHz
2 to 10 kHz
10 to 30 kHz
0.5 + 3
2.0 + 3
4.0 + 10
320.0 mV
3.200 V
32.00 V
320.0 V
1000 V
1.0 + 3
3.0 + 3
N/A
Range
Accuracy
0.3 + 2
320.0 Ω
3.200 kΩ
0.2 + 1
32.00 kΩ
320.0 kΩ
3.200 MΩ
1.0 + 1
32.00 MΩ
32.00 nS
2.0 + 10
Dc current
Range: 0 to 10 A
Accuracy: 0.75 + 2
Ac current
Range: 0 to 10 A
Accuracy: 1.5 + 2