TB 9-6625-2274-50
g. At Command? prompt enter "caldate mm-dd-yy." (For mm-dd-yy, enter actual due
date.)
h. At Command? prompt enter "caldate?" to verify the new caldate.
i. At Command? prompt press Enter to obtain the Instrument Address? prompt.
j.
At Instrument Address? prompt, enter "exit."
k. Repeat a above.
ADJUSTMENT PROCESS FOR
TEKTRONIX, TYPE CG-5011
15. Preliminary
Instructions
a. This section should only be performed if the TI fails to meet the performance
requirements of Section III, Calibration Process for Programmable Calibration Generator,
Tektronix, Type CG-5011.
b. The
instructions outlined in
paragraphs 15 and
16 are preparatory to the
adjustment process. Personnel should become familiar with the entire bulletin before
beginning the calibration. After the adjustment process has been completed, repeat Section
III, Calibration Process for Programmable Calibration Generator, Tektronix, Type CG-
5011.
c. Items of equipment used in this procedure are referenced within the text by common
d. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
personnel fail to observe safety precautions. REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
WARNING
Dangerous potentials exist at several points throughout the TI.
When the instrument is operated with the covers removed, do
not touch exposed connections or components.
a. Remove TI top cover, both side covers, and rear panel. Connect TI to power module
interface using three flexible plug-in extenders.