TB 9-6625-2316-24
5. Accessories Required. The accessories required for the calibration are common usage
accessories, issued as indicated in paragraph 4 above, and are not listed in this calibration
procedure. The following peculiar accessories are also required for this calibration:
Extender Card, part number 385E302; Electric Connector Assembly, part number
M55074/4 (NSN 5935-01-142-9742).
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
(part number)
Common name
Minimum use specifications
Fluke, Model 8840A/AF05
Range: Dc voltage: (30 mV to 3.0 V)
(AN/GSM-64D)
Accuracy: 0.5%
Range : Observe 1000 Hz signal
Agilent, OS-303/G (OS-303/G)
Accuracy: 0.5%
A
Biddle-Gray, Model 71-650
Range: 16.78 k
STANDARD NO. 1
(71-650)
ccuracy: 0.5%
Range: 19.6 to 700 k
A
Biddle-Gray, Model 71-631
STANDARD NO. 2
(7910328)
ccuracy: 0.5%
A
Beckman, Model CR10M
Range: 9 and 10 M
STANDARD NO. 3
(8598965)
ccuracy: 0.75%
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Additional maintenance information is contained in the TM 11-6625-3292-40 and the
manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.