TB 9-6625-2328-24
b. Press TI POWER ON/STBY pushbutton to ON and allow 30 minutes for warm-up.
NOTE
Initialization of TI controls and indicators is accomplished by
cycling the POWER ON/STBY pushbutton from STBY to ON.
a. Performance Check
(1) Press and release TI POWER ON/STBY pushbutton to STBY.
(2) Press TI POWER ON/STBY pushbutton to ON.
(3) Connect pulse generator out1 OUTPUT to TI CHANNEL 1 input.
(4) Press TI keys as listed in (a) through (k) below:
(a) MEASURE Time & Period until TI 1 TO 2 is displayed.
(b) CHANNEL 1 Trigger/Sensitivity until AUTO TRG: is displayed, then
to display AUTO TRG: OFF.
(c) CHANNEL 1 Trigger/Sensitivity until LEVEL: is displayed, use , , ⇐,
to display LEVEL: .500 V and then Enter.
(d) CHANNEL 1 Trigger/Sensitivity until COMMON 1: is displayed, use
to display COMMON 1: ON.
(e) CHANNEL 1 50Ω/1MΩ to 50Ω (LED on).
(f) CHANNEL 1 DC/AC to DC (LED on).
(g) CHANNEL 2 Trigger/Sensitivity until AUTO TRG: is displayed, then
to display AUTO TRG: OFF.
(h) CHANNEL 2 Trigger/Sensitivity until LEVEL: is displayed, use , , ⇐,
to display LEVEL: .500 V and then Enter.
(i) CHANNEL 2 Trigger/Sensitivity until SLOPE: is displayed, then to
display SLOPE: NEG.
(j) CHANNEL 2 50Ω/1MΩ to 50Ω (LED on).
(k) CHANNEL 2 DC/AC to DC (LED on).
(6)
Set pulse generator output as listed in (a) through (h) below:
(a)
(b)
Press the Pulse button, then the Width button, and set to 150.0ns.
(c)
Press the more key to Ampl and set the amplitude for 1V rms output.
(d)
Delay to 0.000.
(e)
Set the Lead edge to 2.5 ns.
(f)
Set the Trail edge to 2.5 ns.
(g)
(h)
Press Out1 button.
(7) Press TI Stop/Single key. If TI does not indicate between 0.146 and 0.154 s,
perform b below.
(8) Disconnect pulse generator from TI.