TB 9-6625-2328-24
(7) Disconnect time/frequency workstation OUTPUT 1 MHz from TI CHANNEL 1 input.
(8) Connect signal generator RF OUTPUT to TI CHANNEL 1 input.
(9) Set signal generator to each output listed in table 6, press TI Stop/Single key as
indicated. TI will indicate within limits specified.
Signal generator
TI
output
indication
TI
(key)
(MHz)
(dBm)
Min
Max
(MHz)
(MHz)
100
-7
Stop/Single
99.999,999,97
100.000,000,03
200
-7
Stop/Single
199.999,999,94
200.000,000,06
225
-7
Stop/Single
224.999,999,93
225.000,000,07
(10)Disconnect signal generator RF OUTPUT from TI CHANNEL 1 input.
(11) Connect time/frequency workstation OUTPUT 1 MHz to TI CHANNEL 2 input.
(12)Press TI MEASURE Freq & Ratio key until TI displays FREQUENCY 2.
(13)Repeat technique of (6) through (10) above for settings in table 7.
Signal generator
TI
output
indication
TI
(key)
(MHz)
(dBm)
Min
Max
(MHz)
(MHz)
100
-7
Stop/Single
99.999,999,97
100.000,000,03
200
-7
Stop/Single
199.999,999,94
200.000,000,06
225
-7
Stop/Single
224.999,999,93
225.000,000,07
(14) Connect signal generator RF OUTPUT to TI CHANNEL 3 input.
(15) Press TI MEASURE Freq & Ratio key until FREQUENCY 3 is displayed.
(16) Set signal generator to each output listed in table 8 and press TI Stop/Single
key as indicated. TI will indicate within limits specified.
Signal generator
TI
output
indication
TI
(key)
Min
Max
(MHz)
(dBm)
(MHz)
(MHz)
100
-17
Stop/Single
99.999,999,97
100.000,000,03
1300
-17
Stop/Single
1,299.999,999,5
1,300.000,000,5
(17)Disconnect equipment from TI.
b. Adjustments. No adjustments can be made.
11