TB 9-6625-2337-24
Table 34. KIR Characteristics - Continued
TI
Test description
Settings/connection
Measurement type
Minimum
Maximum
GTC TRIG RT
--------------------------------------
Rise Time
---------------
100 ns
Channel 2
GTC TRIG FT
--------------------------------------
Fall Time
---------------
200 ns
Channel 2
GTC TRIG
--------------------------------------
Overshoot
---------------
5
OVERSHOOT
Channel 2
GTC TRIG 0.8 μS
--------------------------------------
Width
600 ns
1.0 μs
PW
Channel 2
GTC TRIG DLY
--------------------------------------
Delta time channel 1
371.5 μs
372.5 μs
to channel 2
Move connection from TI KIR
ISLS TRIG 4.5V
V Top
3.5 V
5.5 V
GTC TRIG OUT to TI KIR
LVL
Channel 2
ISLS TRIG OUT.
ISLS TRIG RT
--------------------------------------
Rise Time
---------------
200 ns
Channel 2
ISLS TRIG FT
--------------------------------------
Fall Time
---------------
200 ns
Channel 2
ISLS TRIG
--------------------------------------
Overshoot
---------------
5%
OVERSHOOT
Channel 2
ISLS TRIG PW
--------------------------------------
Width
100 ns
1.5 μs
Channel 2
CHVID - ISLS
Move the connection from the TI
Delta time channel 1
5.8 μs
6.0 μs
TRIGGERS M4 PRE OUT to
to channel 2
DLY
the KIR CH VID OUT.
(6) Reduce all outputs to minimum and disconnect equipment setup.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect as follows:
(a) Connect TI VAR PULSE OUTPUT to pulse generator Ext Input using a
75 feedthrough termination.
(b) Connect pulse generator (9211) OUTPUT to TI KIT EN TRIG IN using a
50 feedthrough termination.
(c) Connect TI IFF VIDEO CHAL/TAG output to KIT CH VID IN using a
75 feedthrough termination.
(d) Connect KIT REPLY OUT to oscilloscope channel 1 input using a 93
feedthrough termination.
(2) Press TI keys as listed in (a) through (g) below:
Press FUNC, ENTR keys.
(a)
Use the arrow keys to highlight GENERAL MENUS and press ENTR key.
(b)
2, ENTR (CHALLENGES).
(c)
Use the arrow keys to highlight M2:.
(d)
ENTR (OFF).
(e)