TB 9-6625-2340-24
Table 2. Minimum Specifications of Equipment Required - Continued
Manufacturer and model
Common name
Minimum use specifications
(part number)
MICROWAVE
Anritsu, Model MF2414B (MF2414B)
FREQUENCY COUNTER
Accuracy: 1 kHz 1 count
Range: 1000 Vdc
Fluke, Model 8840A/AF05
Accuracy: 2 Vdc
(AN/GSM 64D)
Range: Vert. 2 Volts / cm
(OS-303/G)
Horiz. 100 ms / cm
POWER SPLITTER
Range: 20 MHz to 18 GHz
Weinschel, Model 1870A (7916839)
Aeroflex, Model 2023B (2023B) or (SG-
1207/U)
Power range: 0 to -15 dBm
Accuracy: 2 dBm
SYNTHESIZED SIGNAL
Anritsu, Model 68369NV (68369NV)
GENERATOR
Power range: 0 to -120 dBm
NO. 1
Accuracy:1.525 dBm
SYNTHESIZED SIGNAL
Anritsu, Model 68369NV (68369NV)
GENERATOR
Power range:
0 to -120 dBm
NO. 2
Accuracy:
1.525 dBm
Hewlett-Packard, Model 3335AOPT001-
SYNTHESIZER/LEVEL
KO6 (MIS-35938)
Amplitude range: -80 to +10 dBm
Accuracy:
0.02 dB per 10 dB
step
Resolution: 1 x 10-10
TIME/FREQUENCY
Datum, Model ET6000-75
(13589305)
WORKSTATION
1Limited
deploy.
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Additional maintenance information is contained in the manufacturer's manual.
d. When indications specified in paragraphs 8 through 27 are not within tolerance,
perform the power supply check prior to making adjustments. After adjustments are made,
repeat paragraphs 8 through 27. Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to TI.