TB 9-6625-2341-24
(51)
Reset calibrator.
(52) Select ohms 100, 1 k, 10 k, 100 k, 1 M, 10 M, range by pressing pushbuttons
sequence (a) through (w) below:
(a) Toggle
to display/highlight (
CAL 11).
or
or
(b) Set calibrator two wire resistance to produce 100 ohms.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Set calibrator to STBY.
(e) Toggle
to display/highlight (
CAL 12).
or
or
(f)
Set calibrator to produce 1 k ohms.
(g) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(h) Set calibrator to STBY.
(i)
Toggle
to display/highlight (
CAL 13).
or
or
(j)
Set calibrator to produce 10 k ohms.
(k) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(l)
Set calibrator to STBY.
(m) Toggle
to display/highlight (
CAL 14).
or
or
(n) Set calibrator to produce COMP OFF 100 k ohms.
(o) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(p) Set calibrator to STBY.
(q) Toggle
to display/highlight (
CAL 15).
or
or
(r) Set calibrator to produce 1 M ohms.
(s) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(t)
Set calibrator to STBY.
(u) Toggle
to display/highlight (
or
or
CAL 16).
(v) Set calibrator to produce 10 M ohms.
(w) Press (F5) READY, then wait till display/highlight READY is not highlighted.
Set calibrator to standby.
(53)
(54)
Disconnect calibrator leads from TI.
(55)
(56)
Set calibrator to produce +12 volts.
(57)
Press (F3) METER.
(58) Store Cal and Generator out by pressing pushbuttons sequence (a) through (p) below:
(a) (F5) CAL STORE.
(b) Displayed is "New Calibration array will be saved in Flash memory. Are you sure?"
(c)
(F1) YES.
37