TB 9-6625-2341-35
(a) Toggle ▲ or ▼ to display/highlight (▲ or ▼CAL 18).
(b) Set calibrator two wire resistance to produce 100 ohms.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(61) Reset calibrator.
(62) Remove/disconnect both banana wires from calibrator OUTPUT HI and LO from
TI, while leaving BNC circuit connected.
(63) Voltage ramp and 1:1 probe for INPUT A by pressing pushbuttons sequence (a)
through (d) below:
(a) Voltage ramp by toggling▲ or ▼ to display/highlight (▲ or ▼CAL 17).
(b) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(c) 1:1 probe for INPUT A by toggling ▲ or ▼ to display/highlight (▲ or ▼ CAL 21).
(d) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(64) Displayed "Calibration has been completed successfully", then press (F1) CLOSE.
(65) Disconnect BNC cable from INPUT A and reconnect it to INPUT B.
(66) Select 1:1 probe for INPUT B by toggling ▲ or ▼to display/highlight (▲ or ▼CAL 22).
(67) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(68) Displayed "Calibration has been completed successfully", then press (F1) CLOSE.
(69) Connect equipment as shown in figure 14 below:
(70) Select 10:1 red probe for INPUT A by toggling▲ or ▼ to display/highlight (▲ or ▼CAL 19).
(71) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(72) Displayed "Calibration has been completed successfully", then press (F1) CLOSE.
35