TB 9-6625-2341-35
(40)
Press (F3) METER to display/highlight METER.
NOTE
(41) Set calibrator to standby.
(42) Remove +12 V dc from calibrator to TI.
(44) Select linearity calibration, zeroing ranges and ohms ranges by pressing
pushbuttons sequence (a) through (f) below:
(a) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 1).
(b) Press (F5) READY, then wait till display/highlight READY is not highlighted.
NOTE
constants are being set. This process can last up to 3 minutes.
(c) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 2).
(d) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(e) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 10).
(f) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(45) Disconnect all shorts from TI.
(46) Select INPUT A, 300 mV range and zero for open input by toggling ▲ or ▼ to
display/highlight (▲ or ▼ CAL 3). Press (F5) READY, then wait till display/highlight
READY is not highlighted.
NOTE
The readings on the display may not show what's being applied.
(48) Select Channel A, 300 mV, 1 V, 3 V, 30 V, 100 V, 300 V, range/gain by pressing
pushbuttons sequence (a) through (x) below:
(a) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 4).
(b) Set calibrator to produce a 300 mV dc signal.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Set calibrator to STBY.
(e) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 23).
(f) Set calibrator to produce a 1 V dc signal.
(g) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(h) Set calibrator to STBY.
(i) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 5).
(j) Set calibrator to produce a 3V DC signal.
(k) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(l) Set calibrator to STBY.
(m) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 6).
32