TB 9-6625-2341-35
been set to original setting then (F1) CLOSE), which should display/highlight RESET
SCOPE.
(b) Select reset meter by pressing pushbuttons sequence (1) through (3) below:
(1) Toggle ▲ or ▼to display/highlight RESET METER.
(2) (F5) SELECT ITEM.
(3) (For original settings, METER mode needs to be selected. Change to
SCOPE mode now? (F1) YES then (F1) CLOSE) or (if all Meter settings have been set to
original setting then (F1) CLOSE), which should display/highlight RESET METER.
(c) Select reset ohms by pressing pushbuttons sequence (1) through (3) below:
(1) Toggle ▲ or ▼ to display/highlight RESET OHM.
(2) (F5) SELECT ITEM.
(3) (For original settings, ohm needs to be selected. Change to ohm mode now
press (F1) YES then (F1) CLOSE) or (if all ohm settings have been set to original setting
then press (F1) CLOSE), which should display/highlight RESET OHM.
(d) Select reset diode by pressing pushbuttons sequence (1) through (3) below:
(1) Toggle ▲or ▼ to display/highlight RESET DIODE.
(2) (F5) SELECT ITEM.
(3) (For original settings, DIODE needs to be selected. Change to DIODE mode
now press (F1) YES then (F1) CLOSE) or (if all ohm settings have been set to original
setting then press (F1) CLOSE), which should display/highlight RESET DIODE.
(e) Select reset Ext.mV by pressing pushbuttons sequence (1) through (6) below:
(1) Toggle ▲ or ▼ to display/highlight EXTERNAL mV.
(2) (F5) SELECT ITEM.
(3) (For original settings, EXT.mV needs to be selected. Change to EXT.mV
mode now press (F1) YES then (F1) CLOSE) or (if all ohm settings have been set to
original setting then press (F1) CLOSE), which should display/highlight RESET EXT.mV.
(4) (F1) CLOSE.
(5) (F1) CLOSE.
(6) SCOPE METER/MAIN MENU to display SCOPE/METER/Ω/EXT.mV and etc.
(3) Press pushbutton BACKLIGHT (sun shaped pushbutton just below SPECIAL
FUNCT key) with AUTO SET together, then release.
(4) Set calibrator to produce + 12 V dc.
(5) Press (F4) CALIBRATE SCOPEMETER.
(6) Displayed are the numbers of calibration memories free. Press (F1) CLOSE to
display CONTRAST, SCOPE, METER, ESCAPE, and CAL STORE.
(7) Press (F2) SCOPE to display/highlight SCOPE and CAL 1.
(8) Set calibrator to standby.
(9) Remove +12 V dc from calibrator to TI.
(10)Connect equipment as shown in figure 9 below:
27