TB 9-6625-2345-35
(8) Press TI keys as listed in (a) through (e) below:
(a) BLUE SHIFT.
(b) REF LEVEL, 5, 0, kHz dBX.
(c) SINGLE SWEEP.
(d) GREEN SHIFT.
(e) PEAK FIND.
(9) TI MKR LEVEL indication will be within limits specified in first row of table
(10) Press TI keys as listed in (a) through (c) below:
(a) BLUE SHIFT.
(b) MKR OFF.
(c) SAVE A (not illuminated).
(11) Repeat technique of (5) through (10) above for remaining synthesizer/level
generator output amplitudes and TI settings in table 15. TI MKR LEVEL indications and
displayed trace will be within limits specified in table 15.
Synthesizer/level
generator
Test instrument
Amplitude
Ref level
Min RF Atten
Mkr level
indication (dB)
(dBm)
(dBm)
(dB)
Min
Max
60
60
0
9.5
10.5
50
50
10
9.5
10.5
40
40
20
9.5
10.5
30
30
30
9.5
10.5
20
20
40
9.5
10.5
10
10
50
9.5
10.5
0
0
60
9.5
10.5
(12) Rotate TI MIN RF ATTEN dB knob to 0 NORM position.
(13) Set all outputs to minimum and disconnect equipment.
b. Adjustments. None
a. Performance Check
(1) Connect TI RF INPUT 50Ω to synthesizer/level generator OUTPUT 50Ω.
of 22 dBm.
(3) Press TI keys as listed in (a) through (k) below:
(a) BLUE SHIFT.
(b) RESET.