TB 9-6625-2365-35
(8) Set synthesized signal generator RF OUTPUT to on. TI will indicate within
(9) Set synthesized signal generator RF OUTPUT to off.
(10) Repeat technique of (5) through (9) above for synthesized signal generator
frequencies and power meter no. 1 indications listed in table 6. TI will indicate within
Table 6. Input 1
Sensitivity (18 GHz to
26.5 GHz)
Synthesized signal
Power meter
Test instrument indication
generator
no. 1
Limits (Hz)
Indication
(GHz)
(dBm)
Min
Max
18
-28.0
17999999999.7
18000000000.3
19
-28.0
18999999999.7
19000000000.3
20
-25.0
19999999999.7
20000000000.3
21
-25.0
20999999999.7
21000000000.3
22
-25.0
21999999999.7
22000000000.3
23
-25.0
22999999999.7
23000000000.3
24
-25.0
23999999999.7
24000000000.3
25
-25.0
24999999999.6
25000000000.4
26.5
-24.96
26499999999.6
26500000000.4
(11) Reduce all outputs to minimum and disconnect equipment setup.
b. Adjustments. None.
11. Burst Mode
Test (600 MHz to
26.5 GHz)
a. Performance Check
(2) Press TI keys as listed in (a) through (f) below.
(a)
Preset.
(b)
Menu Burst.
(c)
Resolution < or > (until Width is selected).
(d)
Enter (Narrow selected).
(e)
Return to Meas.
(f)
Resolution < or > (to set resolution to 1 MHz).
(3) Adjust synthesized signal generator frequency to 600 MHz and RF output
amplitude controls for a -33 dBm indication on power meter no. 1.
(4) Press TI keys as listed in (a) through (i) below.
(a)
Menu Freq.
(b)
Resolution < or > (until Mode is selected).
(c)
Enter (Manual selected).
(d)
Resolution < or > (until Count is selected).
(e)
Enter (Normal selected).
(f)
Resolution < or > (until Set Freq is selected).
CHANGE 2 11