TB 9-6695-298-35
5. Accessories Required. The
accessories required for this
calibration are
common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
John Fluke, Model 8506A/CT
Voltage range: 0 to +20.00 V dc
(p/o MIS-35947)
Accuracy: 0.001 V dc
Range: -100% to +100%FS
STRAIN GAGE
(P/O MIS-38934 Type 1 and Type 2)
SIMULATOR
Accuracy: 0.0025% of FS
6. Preliminary Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration process. Personnel should become familiar with the entire section before
beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained the manufacturer's manual for this TI.
d. When
indications specified in
paragraph 7 are not within
tolerance, perform the
power supply check prior to making adjustments. After adjustments are made, repeat
paragraph 7. Do not
perform power supply
check if all other
parameters are within
tolerance.
e. Unless otherwise specified, all controls and control settings in this section refer
to the TI.