Figure 4. Error amplifier board - component locations.
9. Program Compensation. a. Performance Check
(4) Verify that 4.1 MHz multiburst packet is of
(1) Adjust waveform monitor (A7) for A-B
equal amplitude to white bar. If not, perform b(l)
differential measurement.
below.
(2) Set manual luminance gain for unity gain
(5) Adjust C2790 and R2990 (fig. 5) for best
through TI.
response of T pulse in sin2 pulse and bar signal.
(3) Adjust manual chroma gain R5854 (fig. 4)
for minimum preshoot on T pulse in sin2 pulse and
bar signal.
8