for 0 ohm.
resistor (meter calibrator) will indicate between 27 and 33
(4) Adjust BATT ADJ control for O-ohm
ohms.
indication on TI meter.
(6) Repeat the procedures given in (3) through
(5) Adjust decade resistor (meter calibrator)
(5) above; use values and control settings listed in table 5.
b. Adjustments. No adjustments can be made.
controls for indication of 30 ohms on TI meter. Decade
Test instrument
decade resistor or meter calibrator
45V BATT
Jumper RX10
indication (ohms)
switch
test lead
meter indication
position
Min
max
OFF
See note below
0
OFF
30
27
33
OFF
RX10
See note below
0
OFF
RX10
30
270
330
OFF
RX100
See note below
OFF
RX100
30
2,700
3,300
OFF
RX1000
see note below
0
OFF
RX1000
30
27,000
33,000
*ON
RX1000
see note below
0
OFF
RX1000
30
27,000
33,000
*ON
RX1000
See note below
0
*ON
RX1000
30
270,000
330,000
*For A and B models, remove the electronic voltmeter lead from jack X and place it in jack - at the top of the TI.
NOTE
Adjust BATTERY ADJUSTMENT control for 0-ohm indication on ohms scale of meter on TI
the TI cannot be adjusted within tolerance, annotate and
10. Final Procedure. a. Deenergize and disconnect all
affix DA Form 2417 (Unserviceable or Limited Use) tag.
equipment and replace TI protective cover.
b. In accordance with TM 38-750, annotate and
By Order of the Secretary of the Army:
FRED C. WEYAND
General, United States Army
Chief of Staff
Official:
PAUL T. SMITH
Major General, United States Army
The Adjutant General
DISTRIBUTION:
To be distributed in accordance with DA Form 12-34A, (qty rqr block no. 75) Requirements for Calibration Procedures
Publications.
U.S. GOVERNMENT PRINTING OFFICE: 1976-665668/519
5