TB 11-6625-2734-50
SECTION IV
head used and may not be properly
NOTE
adjusted for other sampling heads.
Unless otherwise specified,
verify the results of each test
and take corrective action
(3) Depress oscilloscope (A3) right vertical
whenever
the
test
mode switch and right vertical trigger source switch.
requirement is not met before
(4) Adjust TI DC OFFSET control to display
continuing
with
the
perform b(1) below.
(5) Adjust TI mV/DIV VARIABLE control for a 5-
8. Memory Gate Width and Gain. a. Performance
Check.
divisions above 0-percent level on oscilloscope crt
display, perform b(2) and (3) below.
(6) Turn TI mV/DIV VARIABLE switch to CAL IN
OUTPUT to sampling head (A7) lower input, using cable
and remove coaxial line from sampling head input and
(B6).
(2) Connect termination (B11) to sampling head
b. Adjustments.
upper input.
(1) Adjust R592 (fig. 2) until pulse is displayed
NOTE
TI memory gain is adjusted to
compensate for gain of sampling
Figure 1. Adjustment locations on right index card.
(2) Adjust R390 (fig. 2) to raise dot above step 0
100 percent level.
percent level to its maximum amplitude (may overshoot
5