TB 11-6625-2734-50
Figure 2. Adjustment locations - vertical and horizontal cards.
(3) Adjust C275 (fig. 2) to place first dot above
(1) Depress oscilloscope (A3) left vertical mode
0-percent level at 90 percent level (4.5 divisions above 0
switch and connect probe (B9) to vertical amplifier (A8)
CH1 input.
(2) Turn vertical amplifier VOLTS/DIV switch to
10 mV and set AC-GND-DC switch to GND.
(3) Adjust vertical amplifier position control to
center trace on oscilloscope crt and set AC-GND-DC
switch to DC.
(4) Connect probe tip to TI OFFSET OUT jack
and adjust DC-OFFSET control to center trace on
oscilloscope crt. Do not move DC OFFSET control
during the remainder of this check.
(5) Connect probe tip to TI VERT SIG OUT
b(1) below.
(6) Depress oscilloscope right vertical mode
switch and turn mV/DIV VARIABLE control throughout
its range. If trace on oscilloscope crt shifts, perform b(2)
below.
(7) Turn TI mV/DIV VARIABLE control to CAL
Figure 3. Waveform for R390 and C275 adjustments.
IN. If trace is not centered on oscilloscope crt, perform
b(3) below.
9. Memory, Variable, and Vertical Position Balance.
b. Adjustments.
a. Performance Check.
(1) Adjust R245 (fig. 2) until trace is centered on
oscilloscope crt.
6