TB 3-6665-314-50
SECTION III
CALIBRATION PROCESS
identification of equipment referenced by item numbers
NOTE
Unless otherwise specified, verify the
7. Equipment Setup (fig. 2). a. Adjust mechanical
results of each test and take
zero-adjustment on the TI PRESSURE gage.
corrective action whenever the test
b. Set POWER switch to OFF.
requirement is not met before
c. Turn BLEED VALVE fully counterclockwise.
continuing with the calibration.
d. Connect power cable connector to 115-volt, 60
Hz receptacle.
6. Preliminary Instructions. a. The instructions
e. Set POWER switch ON. Motor should operate.
outlined in this section are preparatory to the calibration
f
Allow TI to operate for at least 10 minutes.
process. Personnel should become familiar with the
entire bulletin before beginning the calibration.
b. Items of equipment used in this procedure are
referred within the text by common name and item
identification number as listed in tables 2 and 3. For the
Figure 2. Locations of TI controls and instruments used during calibration.
3