TB 9-1430-201-50
a. Personnel should become familiar with
sections I, II, and IV
before beginning the
b. Items of equipment used in this procedure are referenced within the text by common
name and item identification number as listed in tables 2 and 3. For the identification of
equipment referenced by item numbers prefixed with A, see table 2, and for prefix B, see
Figure 1. Ratio accuracy (Beckman Models) - equipment setup.
11. Input Resistance
a. Performance Check
(1) Connect resistance bridge (A2) UNKNOWN terminals between TI IN and COM
terminals, using Kelvin clips (B4).
(2) Measure TI input resistance. Resistance will indicate within 0.05 percent of
input resistance specified for TI.
a. Adjustments. No adjustments can be made.
7