TB 9-1430-201-50
Test instrument
model
Min
Max
T10AR100
99Ω
101Ω
T10AR500
495Ω
505Ω
T10AR1k
kΩ
kΩ
0.99
1.01
T10AR5k
kΩ
kΩ
4.95
5.05
T10AR10k
kΩ
kΩ
9.9
10.1
T10AR20k
kΩ
kΩ
19.8
20.2
(TA-1001)
T10AR50k
kΩ
kΩ
49.5
50.5
T10AR100k
kΩ
kΩ
99
101
b. Adjustments. No adjustments can be made.
8. Ratio Accuracy
a. Performance Check
(2) Position TI outer dial to 10 and inner dial to 0.
(3) Adjust voltage divider (A3) to 9900000.
(4) Adjust output of voltage standard (A4) to 5 volts.
(5) Set TI outer dial to 9 and inner dial to 90.
(6) Adjust voltage divider dials to obtain null on null detector (A1). Voltage divider
will indicate between 98900 and 99100. If not, perform b below.
(7) Repeat technique of (5) and (6) above at TI dial settings listed in table 5.
Voltage divider will indicate within limits specified. If not, perform b. below to obtain best
in-tolerance compromise of error across entire TI range.
Table 5. Ratio Accuracy
Test instrument
Voltage divider
Voltage divider final indication
Outer dial setting
Inner dial setting
initial setting
Min
Max
9
80
98000
97900
98100
9
70
97000
96900
97100
9
60
96000
95900
96100
9
50
95000
94900
95100
9
40
94000
93900
94100
9
30
93000
92900
93100
9
20
92000
91900
92100
9
10
91000
90900
91100
5