TB 9-4910-556-35
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance checks where applicable.
a.
Connect equipment as
shown in CONNECTION A
(fig. 8).
Figure 8. 10 V
and 100 mV
channel accuracy -
equipment accuracy.
b. Set calibrator to STANDBY and 24 V output.
c. Set CIB PWR switch to ON.
d. Press VTM power switch to ON.
e. Set TEST SELECT switches on VTM to 00.
NOTE
pushbutton on lower left of set communicator.