TB 9-4910-556-35
(2) Adjust controls of calibrator for 3 V, 400 Hz output.
(3) Using DATA ENTRY keys, enter 66658 on set communicator and press GO
key. Set communicator will indicate between 380 and 420.
b. Adjustments. No adjustments can be made.
SECTION V
SET COMMUNICATOR (SC)
NOTE
24. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 24 and
25 are preparatory to the
calibration
process. Personnel should become familiar with the entire section before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in tables 2 and 3.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in TM 9-4910-751-14&P for this TI.
d. Unless otherwise specified, all controls and control settings refer to the set
communicator (SC).
WARNING
HIGH VOLTAGE is used or exposed during the
may result if personnel fail to observe safety precautions.
REDUCE OUTPUT(S) to minimum after each step within
the performance check where applicable.
a. Connect cable W2 to set communicator J2 connector and red and black leads to
calibrator output.
b. Connect set communicator to VTM J2 connector, using cable CX309 (or CX50 and
CX309).