TB 9-4920-364-24
(9) Repeat technique of (4) through (7) above for each setting of RESISTANCE
CHECK SW-3 switch and set resistance standard to the value indicated on SW-3. TI
GALVO-1 will indicate within 5 divisions at all SW-3 settings.
b. Adjustments. Adjust RES CHECK TRIMPOT (fig. 2) for 0 indication on GALVO-1 (R).
10. Insulation Resistance Check
a. Performance Check
(2) Connect TI INPUT RES. INSUL. P-2 to HIGH and LOW terminals of
resistance standard using cables and insulation check adapter (BH485 and BH821, p/o TI).
(3) Set SELECTOR SWITCH SW-1 to INSUL. and SW-8 switch to RX100. If
INSULATION CHECK meter does not indicate 0, perform b (1) below.
(4) Adjust resistance standard for indication of 10,000 ohms on INSULATION
CHECK meter. If resistance standard does not indicate between 9000 and 11,000 ohms,
perform b (2) below.
(6) Set SW-8 switch to RX1000. If INSULATION CHECK meter does not indicate
zero, perform b (3) below.
(7) Adjust resistance standard for an indication of 100,000 ohms on INSULATION
CHECK meter. If resistance standard does not indicate between 90,000 and 110,000 ohms,
perform b (4) below.