TB 9-4920-453-35
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be
used in this
calibration
procedure. This equipment is issued with Secondary Transfer Calibration Standards Set AN/GSM-
287. Alternate items may be used by the calibrating activity when the equipment listed in table 2 is not
available. The items selected must be verified to perform satisfactorily prior to use and must bear
evidence of current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the standard and
TI.
5. Accessories Required. Accessories required for this calibration
are issued as
indicated in
Table 2. Minimum Specifications of
Equipment Required
Minimum use
Manufacturer and model
Item
Common name
specifications
(part number)
A1
AC CALIBRATOR
Range: 50 to 1000 Hz
Hewlett-Packard, Model
0 to 800 mV ac
745AOPTC93 (745AOPTC93)
Accuracy: 1%
w/HV amplifier C90-746A
(C90-746A)
A2
General Radio, Model
Accuracy: 1%
W10MT3AS3 or Ridge,
Model 9020F (7910809)
SECTION III
6. Preliminary Instructions
a. The instructions outlined in
paragraphs 6 and 7 are preparatory to
the calibration process.
Personnel should become familiar with the entire bulletin before beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by common name and
item identification number as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test requirement is
not met, take corrective action before continuing with the calibration. Adjustments required to calibrate
the TI are included in this procedure. Additional maintenance information is contained in the
manufacturer's manual and TM 55-4920-243-15 for this TI.