Kit Simulator Test - Continued
Figure 3. Counter/Decoder - Equipment Setup.

Calibration Procedure For Transponder Test Set An/Apm-421 Page Navigation
  14    15    16    17    18  19  20    21    22    23    24  
img
. . .
TB 9-4920-456-24
(16) On radar test set, press pushbutton S5: ON (highlighted). SLS indicator light
will go on.
(17) On radar test set, press pushbutton M4: DISP to NORM (highlighted) and M4:
OFF (highlighted).
(18) Press TI CODE SEL key on MODE 4 keyboard to select word A (B light off,
A light on).
(19) Set TI switches on MODE 4 panel to OUT position as listed in (a) through (d)
below:
(a) VIDEO (located above VIDEO jack on MODE 4 panel).
(b) TRIG (located above TRIG jack on MODE 4 panel).
(c) REPLY (located above REPLY jack on MODE 4 panel).
(d) DISPARITY (located above DISPARITY jack on MODE 4 panel).
b. Adjustments. No adjustments can be made.
11. Counter/Decoder Test
a. Performance Check
(1) On radar test set, press pushbutton FUNC, ENTR. Wait a second and press
ENTR pushbutton and then ENTR pushbutton again.
(2) Connect equipment as shown in figure 3.
(3) On radar test set, press pushbuttons as indicated in (a) through (e) below:
(a) Starting  from  MENU  SELECTIONS  (highlighted),  select
menu
CHALLENGES (highlighted) and press pushbuttons as listed in 1 through 3 below:
M1: ON.
1
2
P1: ON.
3
P3: ON.
20