SECTION IV. CALIBRATION PROCESS FOR AC GENERATOR DETECTOR, MODEL 860A
Figure 3. Ac detector sensitivity and selectivity - equipment setup.

Calibration Procedure For Model 230B, 860A, 865A, 290A, B, 801, And Sp2979 Page Navigation
  8    9    10    11    12  13  14    15    16    17    18  
img
. . .
TB 9-4931-217-40
(2) Connect lead from chassis to one end of R20 (fig. 2).
Figure 2. Ac generator detector - top interior view.
(3) Connect equipment as shown in figure 3.
(4) Turn GEN VOLTAGE control cw out of OFF position. Allow at least 5 minutes
for equipment to warm up and stabilize.
(5) Rotate DET GAIN control fully ccw.
(6) Adjust function/arbitrary generator for a 200 mV rms (millivolts root-mean-
square) output with frequency at 1 kHz (kilohertz) and output impedance for 50 ohms.
(7) Move multimeter and cable from DET INPUT terminals to DET OUTPUT
terminals.
(8) Fine tune function/arbitrary generator frequency control for peak indication on
multimeter. Frequency counter will indicate between 990 and 1010 Hz (hertz).
(9) Adjust DET GAIN control between fully ccw and approximately one-fourth turn
cw for a convenient multimeter indication. Record multimeter indication.
(10) Increase function/arbitrary generator frequency until multimeter indication is -
30 dB (decibel) from indication recorded in (9) above. Frequency counter indication will be
2020 Hz or less.
13