Table 10. D-Q Dial Linearity
Final Procedure

Calibration Procedure For Model 230B, 860A, 865A, 290A, B, 801, And Sp2979 Page Navigation
  18    19    20    21    22  23  24    25    26    27    28  
img
. . .
TB 9-4931-217-40
Table 10. D-Q Dial Linearity- Continued
Test instrument D-Q dial indications
Test instrument
Resistance standard
D-Q switch
Indications
settings
(ohms)
Min
Max
.01X
1114
69
71
.01X
1273
79
81
.01X
1432
89
91
.01X
1592
99
101
.1X
1.59 k
9
11
.1X
3.18 k
19
21
.1X
4.77 k
29
31
.1X
6.37 k
39
41
.1X
7.96 k
49
51
.1X
9.55 k
59
61
.1X
11.14 k
69
71
.1X
12.73 k
79
81
.1X
14.32 k
89
91
.1X
15.92 k
99
101
b. Adjustments. No adjustments can be made.
22. Series Inductance
a. Performance Check
(1) Position controls as listed in (a) through (c) below:
(a) FUNCTION switch to SERIES L.
(b) RANGE switch to 10-1 H.
(c) MAIN DIAL to 1.0000.
(2) Connect fixed inductance standard to TI UNKNOWN terminals 1 and 2.
(3) Adjust ac generator detector DET GAIN control for a convenient indication on
null meter.
(4) Adjust MAIN DIAL and D-Q dial for null indication on ac generator detector
null meter.
(5) Repeat (3) and (4) above for best null indication on null meter.
(6) Convert MAIN DIAL and RANGE switch indications to inductance. Results
obtained will be within (0.1% +1 dial division) of induction test report value.
(7) Rotate ac generator detector DET GAIN control fully ccw.
b. Adjustments. No adjustments can be made.
23. Parallel Inductance
a. Performance Check
(1) Set D-Q multiplier switch to .01X and FUNCTION switch to L PARALLEL.